Prepare the Inclinometers
-
- Connect the Main Unit to the Companion Unit with the connecting cable.
- Press the START button on the Main Unit.
- Ask individual to place head in neutral position.
- Place the Main Unit on the top of the head.
- Place the Companion Unit over the skin covering the T1 spinous process.
- Press ZERO button to set the neutral position.
|
-
Take Mesurements
-
Cervical Flexion:
-
- Ask individual to flex head forward completely.
- When individual reaches maximum flexion, press the HOLD button to store the maximum flexion angle (a - b).
- Ask individual to return to the neutral position.
- Repeat measurements from three to six times to achieve validity crideria of ±5° or 10%.
-
Cervical Extension:
-
- For cervical extension measurements, the above instructions may be repeated starting with the neutral position. and asking the individual to extend the head backward instead.
|
-
View Data
-
- To view stored data, press the VIEW button.
- Pressing the VIEW button repeatedly will display stored readings, minimum, maximum, and average values.
|
Important Positions |
|
|
|